High Energy Resolution Triple-Axis Spectrometer
C1-1 HER


../../PP/GIF/HER.gif
[Japanese]

Overview:

¡¡A triple axis spectrometer HER is located on the beam port of C1-1 on the cold guide C1 in the guide hall. Inelastic neutron scattering experiments with high-energy resolution can be performed by long wavelength neutrons obtained by the cold guide. The spectrometer consists of three tables: the sample table, analyzer table, and detector table, which are driven by air floating scheme. Energy resolution is in the range of 0.02 to 0.5 meV, depending on incident wavelength and collimation. Energy transfer is in the range of 0.1 to 9 meV. A combination of cooled Be (cutoff energy of 5 meV) and PG Bragg reflection filters are also available to be placed before sample. The spectrometer is controlled by the software named FILMAN-J on Linux.

Specifications:

MonochromatorPG002 vertically focused, 120 mm (H) x 70 mm (V)
45 ° ¡ã 2 θM ¡ã 130 °
(PG: 2.2 meV < Ei < 12meV)
Collimator 1st
2nd 10',20',Open
3rd 20',40',80', 3 Radial Colls for HFA
4th 20',40',80'
FilterCTI cooled Be Filter can be chosen, depending on incident wavelength
rotating PG Bragg reflection filter
Sample heavy-load-proof goniometer
-30 °¡ã 2 θS ¡ã 130 °
Analyzer
horizontally focused (to be installed)
-10 °< 2 θA < 115 °
Detector

horizontally 3He gas detector ¡Êφ 2 inch, effective length 100 mm¡Ë

Beam size25 mm (H) x 40 mm (V)
Control softwareFILMAN-J (PC, Linux version) (temperature controlled using TEMCON)
can be connected by telnet



Contact person:

Dr. Naofumi ASO
NSL-ISSP, Univ. of Tokyo